●High Temperature Fatigue Test System
High-temperature fatigue strength test can be conducted by this system, which controls cyclic loads temperature condition with inert gas.
EMH250 High Temperature Fatigue Test System
Constant temperature chamber
Temperature range |
-30 to +150°C |
-50 to +150°C |
Temperature distribution |
±3°C(empty room) |
Heating time |
90min(RT to +150°C) |
Cooling time |
120min(RT to -30°C) |
Refrigerator |
One-dimensional refrigeration system |
Two-dimensional refrigeration system |
Internal dimensions of bath |
W340×D460×H550mm |
Power supply |
3-phase200V 8kVA |
3-phase200V 10kVA |
Cooling water |
30L/min at 30°C |
45L/min at 30°C |
*Applicable over a range of 60~ max. 450°C in case of high-temperature chamber
EMH250 High Temperature Fatigue Test System
Furnace
Temperature range |
300 to 700°C |
300 to 1500°C |
Temperature distribution |
±3°C(empty furnace) |
External dimensions |
φ50×H500mm |
φ335×H460mm |
Structure |
Half-split annular furnace |
Heating system |
Nichrome wire heater |
Molybdenum disilicide heater |
Power supply |
Single-phase 200V, 6kVA |
Cooling water |
10L/min at 30°C |
|